Aroclor 1221 PCBs on Rtx-CLPesticides

Peaks
1.Decachlorobiphenyl (DCB)
image
GC_EV00993
ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
SampleAroclor 1221 in hexane diluted to 1,000 ppb
decachlorobiphenyl (BZ #209) in acetone diluted to 100 ppb (cat.# 32029)
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.3 min)
Liner:Cyclo Double Gooseneck (4mm) (cat.# 20895)
Inj. Temp.:250 °C
Pulse Pressure:30 psi (206.8kPa)
Oven
Oven Temp.:120 °C to 200 °C at 45 °C/min to 230 °C at 15 °C/min to 330 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant linear velocity
Linear Velocity:60 cm/sec @ 120 °C
DetectorECD @ 330 °C
NotesThis chromatogram was obtained using an Agilent μ-ECD. To obtain comparable results, you will need to employ a μ-ECD in addition to confirmational dual-columns connected to a 5-meter guard column using a “Y” Press-Tight connector.

RECHERCHES SIMILAIRES

aroclor

,

pcb

,

pcbs

,

dcb

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.


Restek Domestic Customer Service

Objet

Votre message

Nom et Prénom

Adresse e-mail

Votre société

Adresse de l'entreprise

Spam Block (Please leave this blank)
Glissez-déposez vos pièces jointes ici (optionnel)
...

Tous les champs sont requis

Merci.

Votre message a bien été envoyé. Nous vous recontacterons dans les plus brefs délais.

Message non envoyé.

Votre message ne peut pas être envoyé pour le moment, veuillez nous excuser pour la gêne occasionnée. Réessayez plus tard ou contactez Restek au 01 60 72 32 10.

www.restek.com/Contact-Us