Used Motor Oil Composite Standard on Rxi-5HT (15 m x 0.25 mm x 0.10 µm)

PeakstR (sec)
1.C40856.7
image
GC_EV1382
ColumnRxi-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905)
SampleUsed motor oil composite standard (cat.# 31465)
Diluent:Methylene chloride
Conc.:50,000 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:87.5 mL/min
Oven
Oven Temp.:40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 420 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesActual column length was calculated to be 15.7 m using holdup time and EZGC flow calculator (http://www.restek.com/ezgc-mtfc).
Voir en PDF

RECHERCHES SIMILAIRES

Total Petroleum Hydrocarbons

,

Total Recoverable Petroleum Hydrocarbons

,

Alkanes

,

Diesel

,

Hump-o-gram

,

Motor Oil

,

Used Motor Oil

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.


Restek Domestic Customer Service

Objet

Votre message

Nom et Prénom

Adresse e-mail

Votre société

Adresse de l'entreprise

Spam Block (Please leave this blank)
Glissez-déposez vos pièces jointes ici (optionnel)
...

Tous les champs sont requis

Merci.

Votre message a bien été envoyé. Nous vous recontacterons dans les plus brefs délais.

Message non envoyé.

Votre message ne peut pas être envoyé pour le moment, veuillez nous excuser pour la gêne occasionnée. Réessayez plus tard ou contactez Restek au 01 60 72 32 10.

www.restek.com/Contact-Us