Resprep CarboPrep Plus SPE Cartridge (Lot 1226 Comparison)

Peaks
1.2,4,5,6-Tetrachloro-m-xylene (TCMX)
Peaks
2.Decachlorobiphenyl (DCB)
image
GC_EV1494
ColumnRtx-CLPesticides, 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
Sample2,4,5,6-Tetrachloro-m-xylene (cat.# 32027)
Decachlorobiphenyl (BZ #209) (cat.# 32029)
Diluent:Hexane
Conc.:2 ppb
Injection
Inj. Vol.:4 µL splitless (hold 0.75 min)
Liner:Topaz 4.0 mm ID single taper inlet liner w/wool (cat.# 23303)
Inj. Temp.:250 °C
Purge Flow:50 mL/min
Oven
Oven Temp.:90 °C (hold 0.5 min) to 330 °C at 25 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:3.5 mL/min
DetectorMicro-ECD @ 340 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC
NotesConditioned a Resprep CarboPrep Plus SPE cartridge (cat.# 25845) by adding one cartridge volume of hexane:acetone (90:10) and letting it stand for five minutes before drawing the solvent down to frit level. Then, 1 mL of sample extract was loaded onto the cartridge and eluted with 9 mL of hexane:acetone (90:10).
Voir en PDF

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.


Restek Domestic Customer Service

Objet

Votre message

Nom et Prénom

Adresse e-mail

Votre société

Adresse de l'entreprise

Spam Block (Please leave this blank)
Glissez-déposez vos pièces jointes ici (optionnel)
...

Tous les champs sont requis

Merci.

Votre message a bien été envoyé. Nous vous recontacterons dans les plus brefs délais.

Message non envoyé.

Votre message ne peut pas être envoyé pour le moment, veuillez nous excuser pour la gêne occasionnée. Réessayez plus tard ou contactez Restek au 01 60 72 32 10.

www.restek.com/Contact-Us