Bleed Comparison of Rxi-624Sil MS and VF-624ms

Peaks
1.Fluorobenzene
image
GC_GN1147
ColumnRxi-624Sil MS (see notes), 30 m, 0.25 mm ID, 1.4 µm (cat.# 13868)
SampleFluorobenzene (cat.# 30030)
Diluent:methanol
Conc.:200 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 20:1)
Liner:4mm Split Liner with Wool (cat.# 20781)
Inj. Temp.:220 °C
Oven
Oven Temp.:40 °C (hold 5 min) to 60 °C at 20 °C/min (hold 5 min) to 120 °C at 20 °C/min (hold 5 min) to 200 °C at 20 °C/min (hold 10 min) to 260 °C at 20 °C/min (hold 10 min) to 300 °C at 20 °C/min (hold 20 min)
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
InstrumentAgilent/HP6890 GC
NotesColumns are of equivalent dimensions and were tested after equivalent conditioning.

RECHERCHES SIMILAIRES

624

,

bleed

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.


Restek Domestic Customer Service

Objet

Votre message

Nom et Prénom

Adresse e-mail

Votre société

Adresse de l'entreprise

Spam Block (Please leave this blank)
Glissez-déposez vos pièces jointes ici (optionnel)
...

Tous les champs sont requis

Merci.

Votre message a bien été envoyé. Nous vous recontacterons dans les plus brefs délais.

Message non envoyé.

Votre message ne peut pas être envoyé pour le moment, veuillez nous excuser pour la gêne occasionnée. Réessayez plus tard ou contactez Restek au 01 60 72 32 10.

www.restek.com/Contact-Us