Permanent Gases on Rt-Msieve 5A (15 m x 0.25 mm x 20 μm)

PeakstR (min)Conc. (Mol%)
1.Helium0.58Balance
2.Argon0.883
3.Oxygen0.912
4.Nitrogen1.745
5.Methane2.344
6.Carbon monoxide7.625
image
GC_PC1349
ColumnRt-Msieve 5A, 15 m, 0.25 mm ID, 20 µm (cat.# 19773)
SamplePermanent gases (custom mix)
Injection
Inj. Vol.:30 µL split (split ratio 40:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:30 °C (hold 10 min)
Carrier GasH2, constant pressure (8.0 psi, 55.2 kPa)
DetectorTCD @ 200 °C
InstrumentAgilent 7890B GC
NotesResolution (Ar/O2) = 1.7
Asymmetry (CO) = 1.2

Note that the sample was introduced onto the column using syringe injection and a high split ratio, which results in a very narrow sample band. Any dead volume in the system (e.g., from using a sampling valve, large sampling loops, or even from using splitless injection) will result in peak broadening, which might decrease resolution between the peaks or even result in peak coelution.
Voir en PDF

RECHERCHES SIMILAIRES

MolSieve

,

Molecular Sieve

,

Natural Gas

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.


Restek Domestic Customer Service

Objet

Votre message

Nom et Prénom

Adresse e-mail

Votre société

Adresse de l'entreprise

Spam Block (Please leave this blank)
Glissez-déposez vos pièces jointes ici (optionnel)
...

Tous les champs sont requis

Merci.

Votre message a bien été envoyé. Nous vous recontacterons dans les plus brefs délais.

Message non envoyé.

Votre message ne peut pas être envoyé pour le moment, veuillez nous excuser pour la gêne occasionnée. Réessayez plus tard ou contactez Restek au 01 60 72 32 10.

www.restek.com/Contact-Us