OVIs by Dynamic Headspace and Gerstel MACH on Rtx-624 (G43) and Rtx-WAX (G16)

Peaks
1.Methanol
2.Acetonitrile
3.Dichloromethane
4.trans-1,2-Dichloroethylene
5.Hexane
6.cis-1,2-Dichloroethylene
7.Nitromethane
8.Tetrahydrofuran
9.Chloroform
10.Cyclohexane
11.1,2-Dimethoxyethane
Peaks
12.Trichloroethylene
13.Methylcyclohexane
14.1,4-Dioxane
15.Pyridine
16.Toluene
17.2-Hexanone
18.Chlorobenzene
19.Ethyl benzene
20.m-Xylene
21.p-Xylene
22.o-Xylene
23.Tetralin
* septum components
image
GC_PH00845
ColumnsRtx-624 20 m, 0.18 mm ID, 1.00 µm (cat.# 40924)
and Rtx-WAX (G16) 20m, 0.18mm ID, 0.4µm (custom)
using IP Deactivated Guard Column 5 m, 0.32 mm ID (cat.# 10044)
with Universal “Y” Press-Tight Connector (cat.# 20405)
SampleResidual Solvents Class 2 - Mix A (cat.# 36271)
Residual Solvents Class 2 - Mix B (cat.# 36272)
Diluent:0.5mL DI water/~0.2g sodium sulfate in 20mL headspace vial; components at 0.10 - 6.00µg
Injectionheadspace-trap split (split ratio 20:1)
Headspace-Trap
Instrument:Tekmar HT3
Inj. Temp.:220 °C
Transfer Line Temp.:220 °C
Valve Oven Temp.:220 °C
Standby flow rate:50 mL/min
Trap Standby Temp:40 °C
Trap Sweep Temp:40 °C
Preheat Mixer:On
Preheat Mixer Time:2.0 min
Preheat Mixing Level:5
Preheat Mixer Stabilize Time:0.5 min
Sample Temp.:80 °C
Sweep Flow Rate:75 mL/min
Sweep Flow Time:5.0 min
Dry Purge: 10.0 min, flow 100 mL/min @ 25 °C
Desorb Preheat Temp:245 °C
Desorb: 1.0 min @ 250 °C
Trap Bake Temp:260 °C
Trap Bake Time:6.0 min
Trap Bake Flow:450 mL/min
Oven
Oven Temp.:250 °C
Carrier GasHe
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
InstrumentAgilent 6890 with Gerstel Modular Accelerated Column Heater (MACH)
NotesSample preheat time: 15.00 min.

Flow rate: Constant Flow, Column 1: 0.85mL/min.; Column 2: 0.99mL/min.

Gerstel temps.:

Column 1: 50°C (2 min.) to 80°C at 20°C/min. (1 min.), to 200°C at 40°C/min. (2 min.)
Column 2: 35°C (2 min.) to 60°C at 100°C/min. (1 min.), to 200°C at 40°C/min. (2 min.)
Voir en PDF

RECHERCHES SIMILAIRES

OVI

,

organic volatile impurity

,

headspace

,

MACH

,

Gerstel

,

residual solvents

,

solvents

,

G16

,

G43

Pour optimiser l'impression de votre chromatogramme, pensez à bien paramétrer l'imprimante !
Dans la fenêtre « Propriétés de l’imprimante » :
1- Choisissez « Ajuster au format de la page »
2- Sélectionner le format « Paysage » pour les chromatogrammes aux nombreux composés,
« Portrait » pour les chromatogrammes aux pics les plus hauts.


Restek Domestic Customer Service

Objet

Votre message

Nom et Prénom

Adresse e-mail

Votre société

Adresse de l'entreprise

Spam Block (Please leave this blank)
Glissez-déposez vos pièces jointes ici (optionnel)
...

Tous les champs sont requis

Merci.

Votre message a bien été envoyé. Nous vous recontacterons dans les plus brefs délais.

Message non envoyé.

Votre message ne peut pas être envoyé pour le moment, veuillez nous excuser pour la gêne occasionnée. Réessayez plus tard ou contactez Restek au 01 60 72 32 10.

www.restek.com/Contact-Us